Key result
Contrast-enhanced C-arm CT correlates strongly with pathology for infarct volume in swine MI models.
Why the study?
Can contrast-enhanced cardiac C-arm computed tomography accurately visualize and quantify the volume of acute and subacute myocardial infarction compared with pathological staining in a porcine model?
Can contrast-enhanced cardiac C-arm computed tomography accurately visualize and quantify the volume of acute and subacute myocardial infarction compared with pathological staining in a porcine model?
Answers are grounded in this paper and cite their sources.
This preclinical study provides proof of concept that contrast-enhanced C-arm CT can quantify acute and subacute myocardial infarct volume directly in the cardiac catheterization laboratory with strong correlation to pathology, though radiation and contrast doses must be reduced before clinical translation.
OBJECTIVES: Cardiac C-arm computed tomography (CT) uses a standard C-arm fluoroscopy system rotating around the patient to provide CT-like images during interventional procedures without moving the patient to a conventional CT scanner. We hypothesized that C-arm CT can be used to visualize and quantify the size of perfusion defects and late enhancement resulting from a myocardial infarction (MI) using contrast-enhanced techniques similar to previous CT and magnetic resonance imaging studies. MATERIALS AND METHODS: A balloon occlusion followed by reperfusion in a coronary artery was used to study acute and subacute MI in 12 swine. Electrocardiographically gated C-arm CT images were acquired the day of infarct creation (n = 6) or 4 weeks after infarct creation (n = 6). The images were acquired immediately after contrast injection, then at 1 minute, and every 5 minutes up to 30 minutes with no additional contrast. The volume of the infarct as measured on C-arm CT was compared against pathology. RESULTS: The volume of acute MI, visualized as a combined region of hyperenhancement with a hypoenhanced core, correlated well with pathologic staining (concordance correlation, 0.89; P < 0.0001; mean [SD] difference, 0.67 [2.98]cm3). The volume of subacute MI, visualized as a region of hyperenhancement, correlated well with pathologic staining at imaging times 5 to 15 minutes after contrast injection (concordance correlation, 0.82; P < 0.001; mean difference, -0.64 [1.94]cm3). CONCLUSIONS: C-arm CT visualization of acute and subacute MI is possible in a porcine model, but improvement in the imaging technique is important before clinical use. Visualization of MI in the catheterization laboratory may be possible and could provide 3-dimensional images for guidance during interventional procedures.
No takes yet. Share an insight, caveat, or question.
Girard et al. (2015) studied Acute and subacute myocardial infarction (porcine model) (n=12). Contrast-enhanced electrocardiographically gated cardiac C-arm computed tomography (CT) vs. Postmortem pathologic staining with 2,3,5-triphenyltetrazolium chloride (TTC) and thioflavin S was evaluated on Acute myocardial infarction volume agreement between C-arm CT and TTC pathologic staining (MD 0.67 cm3 (concordance correlation 0.89), 95% CI 0.78-0.94, p=<0.0001). Contrast-enhanced C-arm CT infarct volume correlated strongly with pathology in acute and subacute MI in swine.
Synapse has enriched 10 related papers. Consider them for comparative context: