Key result
In an adult porcine model, a novel catheter torque tool subjectively improved maneuverability and stability without significantly altering electrical measurements of catheter integrity.
Why the study?
Electrophysiology procedures require manual dexterity to manipulate catheters, and the in-vivo effect of a novel catheter torque tool on catheter integrity remained to be evaluated.
Does the use of a novel catheter torque tool (Peritorq) affect catheter integrity during electrophysiology procedures in an adult porcine model?
Does the use of a novel catheter torque tool (Peritorq) affect catheter integrity during electrophysiology procedures in an adult porcine model?
The Peritorq catheter torque tool improves subjective maneuverability without compromising the electrical or physical integrity of electrophysiology catheters in a porcine model.
Answers are grounded in this paper and cite their sources.
Supports further human evaluation of the torque tool; does not yet justify clinical adoption based on porcine data alone.
INTRODUCTION: Electrophysiology studies and ablation procedures require strength, steadiness, and dexterity to manipulate catheters. We have previously described a novel catheter torque tool (Peritorq) that improves torqueability and stability and decreases user muscle fatigue. The objective was to evaluate measures of catheter integrity with and without the torque tool in place using multiple diagnostic and ablation catheters in an adult porcine model. METHODS: Diagnostic and ablation catheters were inserted through the femoral or jugular vein into areas of the right atrium, coronary sinus (CS), and right ventricle. Electrical measurements including impedance, sensing, and capture thresholds were obtained with and without the torque tool. Ablation lesions (30 s) were given at different locations using both irrigated and nonirrigated catheters and measurements were recorded with and without the torque tool. RESULTS: Procedures were performed in eight adult pigs. Measurements with and without the torque tool in all locations did not differ significantly using any of the catheters. With the nonirrigated ablation catheter there was a significant difference in maximum (mean 1.7 W, p = .03) and average power (mean 9.1 W, p = .04) delivery at the PS tricuspid valve, but there were no other differences with the irrigated or nonirrigated catheters. Subjective assessment by the operator revealed a substantial improvement in maneuverability, ability to transfer torque, and stability within the cardiac space. CONCLUSION: In an in-vivo environment, a novel catheter torque tool subjectively improved catheter manipulation and did not have a significant impact on the integrity of electrophysiologic catheters. Further study including additional catheters and in-vivo human testing is indicated.
No takes yet. Share an insight, caveat, or question.
Mass et al. (2023) studied Electrophysiology studies and ablation procedures (n=8). Novel catheter torque tool (Peritorq) vs. Without the torque tool was evaluated on Electrical measurements including impedance, sensing, and capture thresholds. In an adult porcine model, a novel catheter torque tool subjectively improved maneuverability and stability without significantly altering electrical measurements of catheter integrity.
Synapse has enriched 10 related papers. Consider them for comparative context: