Key result
Electromagnetic interference from an electric razor caused an erroneous discharge of an implantable cardioverter defibrillator due to a defect in the electrode's insulation.
Case Report (n=1)
Electromagnetic interference from household appliances like electric razors can cause inappropriate ICD shocks in the presence of an underlying electrode insulation defect.
Answers are grounded in this paper and cite their sources.
May warrant device interrogation after unexplained shocks near electric razors; leaves open the prevalence of insulation defects enabling household EMI.
We report an unusual case of the erroneous discharge of a third-generation multiprogrammable implantable cardioverter defibrillator in a 64-year-old patient with a history of recurrent ventricular tachycardias caused by electromagnetic interference while shaving with an electric razor. Electromagnetic interference was related to a defect in the electrode's insulation and could not be provoked in an intact electrode.
No takes yet. Share an insight, caveat, or question.
Seifert‐Held et al. (1995) conducted a case report in Recurrent ventricular tachycardias with an implantable cardioverter defibrillator (n=1). Electromagnetic interference from an electric razor was evaluated on Erroneous discharge of implantable cardioverter defibrillator. Electromagnetic interference from an electric razor caused an erroneous discharge of an implantable cardioverter defibrillator due to a defect in the electrode's insulation.