Key result
In a canine model, there was no significant relationship between chronic surface irregularities on explanted polyurethane pacemaker leads and their physical integrity or strength at explant.
Why the study?
Do surface irregularities on explanted polyurethane pacemaker leads correlate with loss of physical integrity or electrical performance in a canine model?
Do surface irregularities on explanted polyurethane pacemaker leads correlate with loss of physical integrity or electrical performance in a canine model?
In a canine model, short-term implantation of polyurethane pacemaker leads resulted in surface irregularities that did not significantly affect physical integrity or electrical thresholds.
Answers are grounded in this paper and cite their sources.
No relationship between surface irregularities and lead integrity in canines; leaves open human applicability and need for prospective validation.
Although over 140,000 polyurethane leads have been implanted in humans, a controversy has recently arisen dealing with the significance of frequently detected irregularities on explanted insulator surfaces by scanning electron microscopy (SEM), and their relationship with lead failure. We therefore implanted ten, 6 Fr atrial polyurethane leads and an equal number of 4 Fr ventricular leads in dogs for a mean 26 +/- 2 and 29 +/- 2 weeks, respectively. Fourteen leads were removed utilizing metered force. The remaining leads were removed surgically without force at the termination of the study. Specimens were stored in saline prior to physical testing (tensile strength and elongation) of intravascular and extravascular lead segments. Similarly stored samples were prepared for SEM analysis. The polyurethane insulator disconnected from the electrode on forcible removal in three atrial and two ventricular leads (concordant connected from the electrode on forcible removal in three atrial and two ventricular leads (concordant for two pairs). There was no significant relationship between the results of physical testing and extraction force, lead size, integrity on removal, sample location or SEM scores. Chemical analysis of surface changes employing both x-ray and electron emission spectroscopy failed to reveal evidence of protein adsorption or liquid impregnation. Samples soaked in sodium hydroxide displayed the same SEM irregularities. The combined effects of fissuring and fluid absorption by polyurethane did not affect chronic anodal or cathodal thresholds. In conclusion, a clear relationship between chronic surface irregularities and the insulator function of polyurethane, its relationship between chronic surface irregularities and the insulator function of polyurethane, its integrity in situ, or its strength at explant has not been established by this relatively short-term study. On the other hand, a potential time-dependent impact of these changes on lead performance has not been excluded. Moreover, only a single species of medical grade polyurethane has been clinically evaluated. Thus, the ultimate place of polyurethane leads in pacing systems has yet to be determined.
No takes yet. Share an insight, caveat, or question.
TIMMIS et al. (1983) studied Pacemaker lead evaluation (n=20). Polyurethane pacemaker leads was evaluated on Relationship between physical testing and extraction force, lead size, integrity on removal, sample location or SEM scores. In a canine model, there was no significant relationship between chronic surface irregularities on explanted polyurethane pacemaker leads and their physical integrity or strength at explant.
Synapse has enriched 10 related papers. Consider them for comparative context: