Why the study?
Do high-stress designs and silver-containing coil wires increase the risk of insulation failure in polyurethane pacemaker leads?
Do high-stress designs and silver-containing coil wires increase the risk of insulation failure in polyurethane pacemaker leads?
Polyurethane pacemaker lead failures are primarily driven by high-stress designs and chemical interactions, highlighting the importance of low-stress designs and nonreactive materials for long-term durability.
Answers are grounded in this paper and cite their sources.
Supports low-stress, silver-free polyurethane leads in selection; leaves open prospective validation of durability.
Environmental stress cracking has been identified as a crack propagating mechanism in polyurethane-insulated, heart pacemaker leads, which is directly related to specific lead design parameters. Lead designs imposing excessive stress on the polyurethane insulation through an interference fit between the coil and polymer have demonstrated insulation failures. Conversely, low-stress designs have shown virtually no insulation problems. The higher-stress designs have used organic solvents to facilitate coil placement during manufacturing, which may result in lowering the polymer's ability to resist the higher stress. In addition, a specific silver-containing coil wire composition has been found to galvanically corrode upon body fluid intrusion into the lead, ionizing the silver. These ions interact with the polyurethane polymer resulting in the loss of polymer strength. All polyurethane lead failures to date have been specific to high stress and/or chemical interaction. Leads using low-stress designs and nonreactive coil wire compositions continue to demonstrate a positive clinical experience.
No takes yet. Share an insight, caveat, or question.
Phillips et al. (1986) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: