The design and performance of a platinum resistance thermometer for high temperatures are described. The thermometer bulb is 36 mm × 5.5 mm and is contained in a recrystallized alumina sheath 500 mm long and 8 mm in external diameter. The coil is wound from very pure platinum wire (α = 0.003 926) of diameter 0.3 mm and is freely exposed to dry air. The coil resistance at 0° C is 1.4 Ω and this permits measurements to the equivalent of 0.002 deg C. The thermometer is found to be stable to this limit on heating it for one hour at 1063° C.