Provides a computational fluid dynamics framework to calibrate orbital shaking tests for evaluating the shear-induced degradation of blood-contacting mechanical heart valve coatings prior to pulsatile testing.
Answers are grounded in this paper and cite their sources.
May guide preclinical shear testing of valve coatings; leaves open pulsatile and in vivo validation before clinical use.
Biomedical coatings are used to promote the wear resistance and the biocompatibility of a mechanical heart valve. An orbital shaking test was proposed to assess the durability of the coatings by the amount material eroded by the surrounding fluid. However, there is still a lack of understanding with regards to the shaker’s rotating conditions and the corresponding physiological condition. This study implemented numerical simulations by establishing a fluid dynamic model to evaluate the intensity of the shear stress under various rotating speeds and diameters of the shaker. The results are valuable to conduct in vitro tests for estimating the performance of biomedical coatings under real hemodynamic conditions and can be applied to other fluid-contact implants.
No takes yet. Share an insight, caveat, or question.
Cherng et al. (2017) studied Degradation of biomedical coatings for mechanical heart valves. Orbital shaking hydrodynamic simulation and in vitro erosion testing (rotating speed 60-300 rpm, rotating diameter 1-5 cm, container inner diameter 2-5 cm) vs. Different layer-by-layer coating configurations (Ti/Dop/(Hep_Col)4/Hep vs Ti/Dop/(Hep_Col)9/Hep) and bare titanium alloy was evaluated on Hydrodynamic shear stress (average and maximum) and remaining coated heparin content.