The accurate determination of E,E-2,4-hexadiendioic acid (t,t-muconic acid), an important urinary biomarker of benzene exposure, is directly performed in human urine by a new two-dimensional liquid chromatographic method. After a first partial separation of urine components by a reversed-phase mechanism and the focalisation of the anionic analytes on a small anion-chromatographic column, a conventional ion-chromatographic analysis with UV detection is carried out. The analytical procedure is fully controlled by the HPLC instrumentation software using an eight-port switch valve. If compared with conventional one-dimensional procedures, the new method produces chromatograms containing a limited number of well resolved peaks and consequently allows better analytical performances: no interfering peaks, absence of bias, repeatability better than 5% in the concentration range 0.09-5 mg dm-3 and a detection limit of 4.0 micrograms dm-3 (alpha = beta = 0.05) for the analysis of real samples.
No takes yet. Share an insight, caveat, or question.
Serena et al. (2000) studied this question.