Key result
Valve-in-valve treatment using a self-expanding TAVR device successfully treated massive aortic regurgitation due to early partial embolization of a Perceval valve.
Case Report (n=1)
This case demonstrates the feasibility of using a self-expanding TAVR device for valve-in-valve treatment of early partial embolization of a sutureless aortic valve.
Answers are grounded in this paper and cite their sources.
May serve as bailout in rare Perceval embolization; leaves open need for prospective safety data.
In the last years, the use of sutureless devices in frail patients with severe aortic stenosis has increased thanks to their "easier and faster" technique of implantation in comparison to conventional surgery. Results from metanalysis show comparable outcomes in comparison to transcatheter aortic valve replacement (TAVR) in terms of mortality, stroke incidence, and rate of pace-maker implantation. The incidence of para-valvular leak (PVL) is even lower for sutureless devices than for TAVR. The few cases described are generally due to incomplete decalcification or incorrect valve sizing and consequent stent distortion. To our knowledge this is the first case describing PVL with massive aortic regurgitation due to early partial embolization of a Perceval valve and its successfully treatment with valve-in-valve by using a self-expanding TAVR device.
No takes yet. Share an insight, caveat, or question.
Laricchia et al. (2019) conducted a case report in Perceval sutureless valve migration with massive aortic regurgitation (n=1). Valve-in-valve treatment using a self-expanding TAVR device (CoreValve Evolut Pro) was evaluated on Successful treatment of para-valvular leak with massive aortic regurgitation. Valve-in-valve treatment using a self-expanding TAVR device successfully treated massive aortic regurgitation due to early partial embolization of a Perceval valve.
Synapse has enriched 10 related papers. Consider them for comparative context: