Key result
Glutaraldehyde fixation of bovine pericardium produces a more extensible material than natural tissue, though tricuspid heterografts show maximal bending stresses at sites associated with fatigue.
Glutaraldehyde fixation of bovine pericardium alters its mechanical properties, identifying specific sites of maximal bending stress that may contribute to bioprosthetic valve failure.
Answers are grounded in this paper and cite their sources.
Fixation alters pericardial mechanics and stress sites; leaves open whether these properties predict heterograft durability in patients.
Glutaraldehyde fixation of bovine pericardium produces a material which is more extensible than the natural tissue. Chemically modified pericardial sacs from 16–20 week old calves provide a region 12 000 mm2 in area which has relatively uniform properties. Tricuspid heterografts constructed from this material have maximal bending stresses at two sites associated with valve failure through mechanical tissue fatigue.
No takes yet. Share an insight, caveat, or question.
Trowbridge et al. (1987) studied this question. Glutaraldehyde fixation of bovine pericardium vs. Natural tissue was evaluated on Extensibility and bending stresses. Glutaraldehyde fixation of bovine pericardium produces a more extensible material than natural tissue, though tricuspid heterografts show maximal bending stresses at sites associated with fatigue.
Synapse has enriched 10 related papers. Consider them for comparative context: