Glutaraldehyde-fixed bovine pericardium can be manufactured to resemble the natural anisotropy of human or porcine heart valve leaflets, supporting its use in bioprostheses.
Answers are grounded in this paper and cite their sources.
Supports anisotropic pericardial bioprosthesis design; leaves open durability and performance in clinical use.
Uniaxial tensile tests were performed on glutaraldehyde fixed bovine pericardial strips prepared from chemically modified pericardial samples. These samples originated from an area which demonstrated anisotropic mechanical properties in the native material and which is suitable for the construction of leaflets for pericardial bioprostheses. After glutaraldehyde fixation the tissue had retained its anisotropicity in stiffness and strength in two orthogonal directions. In the range of the functional stresses for a heart valve leaflet (< 1 MPa) the unconstrained fixation regime had modified the initial anisotropic elastic behavior into a more isotropic one. The implications of these findings are that leaflets manufactured from bovine pericardium can be made to resemble, to a degree, the well known anisotropy found in two orthogonal directions in natural human heart valve leaflets, or porcine bioprosthetic heart valve leaflets.
No takes yet. Share an insight, caveat, or question.
Zioupos et al. (1994) studied this question.