Key result
Delayed defibrillation testing 2 months after CRT-D implantation was effective at first shock in 96% of patients, with a 1% acute LV lead dislodgment rate and no ICD therapy failures.
Why the study?
Is delayed defibrillation testing at 2 months safe and feasible in heart failure patients implanted with CRT-D?
Population
211 consecutive heart failure patients undergoing CRT-D implantation
Comparison
Delayed defibrillation testing performed at 2… vs Defibrillation testing performed at the end of…
Design
Cohort
Follow-up
2 months
Authors
Loading...
Should not yet change practice for defibrillation testing timing after CRT-D; leaves open need for randomized confirmation of safety.
Cohort (n=211)
Is delayed defibrillation testing at 2 months safe and feasible in heart failure patients implanted with CRT-D?
Delayed defibrillation testing 2 months after CRT-D implantation is safe and feasible, demonstrating a high rate of effective first shocks and low complication rates.
Gasparini et al. (2005) conducted a cohort in Heart failure patients treated with CRT-D (n=211). Delayed defibrillation testing (2 months post-implantation) vs. Immediate defibrillation testing (at end of implantation) was evaluated on First shock effectiveness at delayed defibrillation testing. Delayed defibrillation testing 2 months after CRT-D implantation was effective at first shock in 96% of patients, with a 1% acute LV lead dislodgment rate and no ICD therapy failures.