Key result
External enhanced counterpulsation (EECP) showed no signs of inappropriate sensing or noise in patients with implanted ICDs, though inappropriate sinus tachycardia occurred when rate responsive mode was on (P=0.001).
Why the study?
Does external enhanced counterpulsation (EECP) cause device interaction in patients with implanted ICDs or CRT-Ds?
Observational (n=21)
Does external enhanced counterpulsation (EECP) cause device interaction in patients with implanted ICDs or CRT-Ds?
EECP appears safe in patients with ICDs and CRT-Ds, though rate-responsive modes may cause inappropriate sinus tachycardia and unipolar sensing may interfere with EECP synchronization.
BACKGROUND: Potential interference between implanted cardiac devices and other medical instruments is an important concern. Therefore, we aimed to investigate the possible device interaction between implantable cardioverter defibrillators (ICDs) and external enhanced counterpulsation (EECP) treatment. METHODS: Twenty-one patients with an implanted ICD or cardiac resynchronization therapy with defibrillator (CRT-D) were enrolled into the study. EECP had applied as two sessions of 5 minutes. Data from device interrogations before and after the first EECP session and during second EECP session were recorded and analyzed for signs of possible device interaction. RESULTS: There was no sign of inappropriate sensing or noise during EECP session. There was no difference regarding electrode impedance, pacing, and sensing values before and after EECP. There was a statistically significant difference regarding heart rates during EECP therapy between rate response off and on modes (68.69 ± 5.92 beats/min and 90.32 ± 11.05 beats/min, respectively P = 0,001). In four patients with CRT-D and unipolar left ventricular pacing, counterpulsation could not be done because of QRS sensing problems. CONCLUSIONS: EECP seems to be a safe treatment modality in patients with implanted ICD and CRT-D devices. It should be kept in mind that in those patients with CRT-D, rate responsive mode is on; inappropriate sinus tachycardia can be seen during EECP therapy. Also in patients with CRT-D using a unipolar sensing mode, problems of QRS complex sensing by the EECP may occur and, therefore, this effects synchronization and success of EECP therapy.
No takes yet. Share an insight, caveat, or question.
Çelik et al. (2013) conducted an observational in Implanted ICD or CRT-D (n=21). External enhanced counterpulsation (EECP) vs. Before EECP / Rate response off was evaluated on Signs of possible device interaction (inappropriate sensing or noise). External enhanced counterpulsation (EECP) showed no signs of inappropriate sensing or noise in patients with implanted ICDs, though inappropriate sinus tachycardia occurred when rate responsive mode was on (P=0.001).