Key result
Precordial pacing models identify ~10 ms as the optimal stimulus duration to minimize pain.
Theoretical modeling suggests that a pulse duration of approximately 10 ms is optimal for precordial cardiac pacing to minimize pain.
Answers are grounded in this paper and cite their sources.
May improve tolerance of precordial pacing; leaves open prospective clinical validation before any practice change.
In precordial pacing with skin-surface electrodes, the goal is to excite the ventricles with minimal stimulation of overlying tissues. A theoretical analysis is presented to identify the relevant factors; the two most important are electrode location and pulse duration. Using the basic law of stimulation, we developed a model which indicates that the optimum pulse duration for the closed-chest pacing pulse is one that is long with respect to the membrane time constant of cardiac muscle. Current-versus-duration curves are presented for pacing and pain, based on experimentally obtained data. For pacing with minimum pain, the optimum stimulus duration was found to be about 10 ms.
No takes yet. Share an insight, caveat, or question.
Geddes et al. (1985) studied Precordial cardiac pacing. Pulse duration for precordial cardiac pacing was evaluated on Optimum stimulus duration for pacing with minimum pain. Theoretical modeling based on experimental data indicates that the optimum stimulus duration for precordial cardiac pacing with minimum pain is approximately 10 ms.