Key result
Left axillary ICD placement is linked to ~26-fold higher odds of ≤5J VF termination versus chest implantation.
Why the study?
Does left axilla ICD implantation reduce defibrillation threshold and myocardial damage compared to traditional left anterior chest implantation?
Cohort (n=70)
Does left axilla ICD implantation reduce defibrillation threshold and myocardial damage compared to traditional left anterior chest implantation?
Implanting the ICD pulse generator in the left axilla significantly lowers the defibrillation threshold compared to the traditional left anterior chest site, which may reduce shock-induced myocardial damage.
Answers are grounded in this paper and cite their sources.
May support axillary ICD placement to lower defibrillation thresholds; leaves open effects on myocardial damage and outcomes.
BACKGROUND: To reduce myocardial damage caused by implantable cardioverter defibrillator (ICD) shock, the left axilla was studied as an alternative pulse generator implantation site, and compared with the traditional implantation site, the left anterior chest. METHODS AND RESULTS: Computer simulation was used to study the defibrillation conduction pattern and estimate the simulated defibrillation threshold (DFT) and myocardial damage when pulse generators were placed in the left axilla and left anterior chest, respectively; pulse generators were also newly implanted in the left axilla (n=30) and anterior chest (n=40) to compare the corresponding DFT. On simulation, when ICD generators were implanted in the left axilla, compared with the left anterior chest, the whole heart may be defibrillated with a lower defibrillation energy (left axilla 6.4 J vs. left anterior chest 12.0 J) and thus the proportion of cardiac myocardial damage may be reduced (2.1 vs. 4.2%). Clinically, ventricular fibrillation was successfully terminated with a defibrillation output ≤5 J in 86.7% (26/30) of the left axillary group, and in 27.5% (11/40) of the left anterior group (P<0.001). CONCLUSIONS: Clinically and theoretically, the left axilla was shown to be an improved ICD implantation site that may reduce DFT and lessen myocardial damage due to shock. Lower DFT also facilitates less myocardial damage, as a result of the lower shock required.
No takes yet. Share an insight, caveat, or question.
Noro et al. (2016) conducted a cohort in Ventricular arrhythmia requiring ICD or CRT-D (n=70). Left axilla ICD implantation vs. Left anterior chest ICD implantation was evaluated on Termination of ventricular fibrillation with a defibrillation output ≤5 J (OR 25.8, p=<0.001). Left axillary ICD implantation increased the rate of successful ventricular fibrillation termination with ≤5 J compared to traditional left anterior chest implantation.