Key result
Continuous-flow LVADs show similar 3-month peak VO2 compared to pulsatile LVADs.
Why the study?
Do continuous-flow LVADs provide similar functional and haemodynamic recovery compared to pulsatile LVADs in patients with end-stage heart failure?
Observational
Do continuous-flow LVADs provide similar functional and haemodynamic recovery compared to pulsatile LVADs in patients with end-stage heart failure?
Absolute Event Rate: 18.3% vs 20.2%
p-value: p=0.09
Continuous-flow LVADs provide comparable functional and haemodynamic recovery, including exercise performance, to pulsatile LVADs at 3 months in patients with end-stage heart failure.
Supports comparable functional recovery with continuous-flow LVADs; leaves open superiority of pulsatile devices pending randomized confirmation.
AIMS: Caused by ageing of the population, better survival from ischaemic heart disease, and improved treatment of chronic heart disease, the incidence of heart failure has increased enormously. Worldwide, left ventricular assist devices (LVADs) are increasingly being used as a bridge or alternative to heart transplantation. In this study, we investigated whether there is difference in functional and haemodynamic recovery after implantation of pulsatile and continuous-flow pumps. METHODS AND RESULTS: We compared laboratory and echocardiographic data and exercise performance in patients with end-stage heart failure, before and 3 months after implantation of pulsatile and continuous-flow LVADs. A significant improvement in all laboratory parameters after implantation of both types of LVADs was seen, as well as a significant decrease in heart rate and LV dimensions, indicating better haemodynamics and cardiac recompensation. This improvement was better for the pulsatile device, probably due to higher plasma levels and higher LV dimensions before implantation. Exercise capacity strongly improved: 3 months after implantation of pulsatile and continuous-flow LVADs, peak VO(2) was 20.2 ± 4.8 vs. 18.3 ± 4.8 mL/kg/min (P = 0.09) (53 ± 12 vs. 49 ± 11% of predicted for age and gender) (P = 0.28). CONCLUSION: Pulsatile and continuous-flow LVADs result in extensive haemodynamic recovery and exercise performance compatible with daily life activities. Exercise performance with continuous-flow LVADs is equal to that with pulsatile devices. This, in combination with improved survival of the newer devices, allows its use as an alternative to heart transplantation in selected patients.
No takes yet. Share an insight, caveat, or question.
Pruijsten et al. (2012) conducted an observational in End-stage heart failure. Continuous-flow left ventricular assist devices (LVADs) vs. Pulsatile left ventricular assist devices was evaluated on Peak VO2 (mL/kg/min) at 3 months (p=0.09). Continuous-flow LVADs resulted in similar exercise capacity (peak VO2) compared to pulsatile LVADs at 3 months post-implantation (18.3 vs. 20.2 mL/kg/min, P=0.09).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: