Key result
SAPIEN 3 exhibits ~3-fold higher diastolic leaflet stress than CoreValve and surgical PERIMOUNT valves.
Why the study?
Data regarding the long-term durability of transcatheter aortic valves are limited, and differences in durability across devices may exist.
Do transcatheter aortic valves (SAPIEN 3, CoreValve) have comparable leaflet stress distribution to surgical bioprosthesis (PERIMOUNT Magna)?
Do transcatheter aortic valves (SAPIEN 3, CoreValve) have comparable leaflet stress distribution to surgical bioprosthesis (PERIMOUNT Magna)?
Biomechanical analysis reveals that the CoreValve has lower peak diastolic leaflet stress than the SAPIEN 3, comparable to a surgical bioprosthesis, which may impact long-term durability.
CoreValve may offer lower leaflet stress than SAPIEN 3 in this animal model; leaves open implications for transcatheter valve durability.
Objective: Transcatheter aortic valves (TAVs) are prone to structural valve deterioration resulting in limited long-term durability. However, data regarding long-term durability of TAVs are limited and there might exist a difference in long-term durability of different TAVs. The aim of this study was to perform a side-by-side comparison of leaflet stress distribution of three most commonly used transcatheter and surgical bioprostheses. Methods: Using integrated experimental methods (i.e., digital image correlation measurements and hemodynamic testing) and numerical methods (i.e. finite element modeling and material optimization), three-dimensional anisotropic mechanical properties of leaflets of 26-mm Edwards SAPIEN 3 and 26-mm Medtronic CoreValve were characterized and compared to that of a 25-mm Carpentier-Edwards (CE) PERIMOUNT Magna. Maximum in-plane principal stress distribution of the leaflets were obtained and compared during systole and diastole. Results: High stress regions were primarily observed in the fixed boundary edges of the three bioprostheses in the systole. During diastole, high stress regions were observed close to the commissures in the Magna and SAPIEN 3 (Figure 1). For CoreValve, however, the maximum stress values were observed in the lower leaflet belly region. In the diastole, the peak of maximum in-plane principal stress was 0.98, 0.96 and 2.95 MPa for the CE PERIMOUNT Magna, CoreValve, and SAPIEN 3, respectively. Conclusions: The present work presents a reliable approach to compare leaflet stress distribution among different bioprostheses. Our results indicated the CoreValve had the lowest peak stress value at the identical pressure value during diastole compared to the PERIMOUNT Magna and SAPIEN 3. Considering the leaflet stress distributions, there might be a difference in the long-term durability of different TAVs.
No takes yet. Share an insight, caveat, or question.
Abbasi et al. (2019) studied Structural valve deterioration. Transcatheter aortic valves (Edwards SAPIEN 3 and Medtronic CoreValve) vs. Surgical Carpentier-Edwards PERIMOUNT Magna bioprosthesis was evaluated on Peak maximum in-plane principal stress during diastole. In diastole, peak maximum in-plane principal stress was 0.96 MPa for CoreValve and 2.95 MPa for SAPIEN 3, compared to 0.98 MPa for the surgical PERIMOUNT Magna bioprosthesis.