Key result
Severe metabolic acidosis lowers the catheter-induced ventricular fibrillation threshold by ~23% in dogs.
Why the study?
The effect of metabolic conditions on fibrillation thresholds induced by intracardiac catheter leakage currents was not well characterized.
Population
Dogs with normal acid-base and cation state
Comparison
Normal acid-base state vs severe metabolic acidosis vs hypoxaemia
Design
Experimental study
Authors
Loading...
May warrant caution with intracardiac devices during acidosis; leaves open human translation.
Lee et al. (1973) studied this question. A 50 Hz electric shock passed through a ventricular catheter has a minimum fibrillation threshold of 130 uA in dogs, which decreases to 100 uA during severe metabolic acidosis.