Key result
Plastic FBG fingertip sensors highly correlate with reference pulse rates within ±5.5 bpm accuracy.
Why the study?
There is a need to develop sensors that can easily measure vital signs such as pulse rate and blood pressure via arterial strain.
Does a plastic FBG sensor accurately measure pulse rate from fingertip pulsation strain in human participants?
Does a plastic FBG sensor accurately measure pulse rate from fingertip pulsation strain in human participants?
Effect estimate: ±5.5 bpm
A novel plastic FBG sensor can measure fingertip pulsation strain to calculate pulse rate with an accuracy of ±5.5 bpm, demonstrating potential for non-invasive vital sign monitoring.
Plastic FBG sensors may enable fingertip arterial strain measurement; leaves open prospective validation before any clinical use.
There is a need to develop sensors that can measure vital signs such as pulse rate and blood pressure easily; such vital signs can be calculated by measuring arterial strain. To this end, a new method that uses a plastic fiber Bragg grating (FBG) sensor to measure pulsation strain at human fingertips is proposed herein. Plastic fiber has a smaller Young’s modulus than that of the glass fiber. Under the same applied pressure, the plastic FBG sensor deforms more than the silica glass FBG sensor, and the Bragg wavelength shift length becomes longer, which results in higher detection sensitivity. The pulsation strain of the fingertip was measured by the proposed method based on this characteristic. The most sensitive signal detection point in the fingertip was the measurement point 10 or 15 mm from the distal interphalangeal joint in the fingertip direction. Strain signals with periodic peaks were measured from five participants. From the strain signal measured with the cuff attached to the arm, the peak disappeared when the cuff moved, and it reappeared when the cuff was opened. Therefore, the measured signal waveform changes based on the blood flow condition. The pulse rate calculated from the peak interval of the measured signal showed high correlation with the reference pulse rate, and the measurement accuracy was ±5.5 bpm. It may be possible to calculate the blood pressure and pulse rate from the measurement signal of the pulsation strain of the fingertip obtained using the plastic FBG sensor.
No takes yet. Share an insight, caveat, or question.
Koyama et al. (2021) studied this question. Plastic fiber Bragg grating (FBG) sensor vs. Reference pulse rate was evaluated on Pulse rate measurement accuracy (±5.5 bpm). Pulse rate calculated from fingertip pulsation strain using a plastic FBG sensor showed high correlation with the reference pulse rate, with a measurement accuracy of ±5.5 bpm.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: