Key result
Ventricular Automatic Capture linked to ~6% longer projected pacemaker battery longevity versus assumed VAC-Off programming.
Why the study?
The effect of Ventricular Automatic Capture technology on device longevity in dual chamber rate responsive pacemakers was evaluated to address its potential benefit.
Does Ventricular Automatic Capture (VAC) technology improve expected device longevity in patients with dual chamber rate responsive pacemakers?
Observational (n=87)
Does Ventricular Automatic Capture (VAC) technology improve expected device longevity in patients with dual chamber rate responsive pacemakers?
Ventricular Automatic Capture technology significantly increases projected pacemaker battery longevity compared to standard fixed-margin programming.
Answers are grounded in this paper and cite their sources.
May support VAC programming for modest battery extension in pacemakers; leaves open prospective confirmation of real-world gains.
Ventricular Automatic Capture (VAC) is thought to increase device longevity by reducing the pacing safety margin. We evaluated the effect of the GUIDANT VAC technology on longevity in dual chamber rate responsive pacemakers. 87 patients, 69.8±13.6 years indicated for DDDR pacing received an INSIGNIA I Ultra DR pacemaker with VAC technology and had a follow-up after 3 months (3 MFU). 29 different lead types from 9 manufacturers were used. Measured ventricular pacing threshold, lead impedance, %A/V-paced, and average pacing rate were used to calculate expected device longevity. Comparisons were made with the assumption that the VAC-Off ventricular amplitude would have been programmed to 2x the 3-MFU measured threshold, but at least 2.5V. Atrial: % pacing: 42±35, lead impedance: 510.7±188.2W, pacing voltage: 2.9±0.9V. Ventricular: % pacing 80±32, lead impedance: 649.9±218.6W, pacing voltage: 1.2±0.3V. Ventricular pacing amplitude using VAC was 1.2±0.3V and without VAC 2.5±0.2V (p<0.01). There was a significant increase in projected device longevity between VAC-On (7.2±0.7 years) and VAC-Off (6.8±0.8 years) (p<0.01). VAC may result in increased battery longevity over a wide range of ventricular leads.
No takes yet. Share an insight, caveat, or question.
Sperzel et al. (2005) conducted an observational in Indicated for DDDR pacing (n=87). Ventricular Automatic Capture (VAC) vs. VAC-Off (assumed programming) was evaluated on Projected device longevity (p=<0.01). Ventricular Automatic Capture increased projected pacemaker battery longevity compared to assumed VAC-Off programming.