Key result
Exchanging a pacemaker generator to one with an impedance-derived contractility sensor (closed-loop stimulation) resolved sudden breathlessness and heart rate drops in a 74-year-old elite cyclist.
Why the study?
Effective rate-adaptive pacing is challenging in patients with atrial fibrillation, especially during high-intensity exercise.
Does an impedance-derived contractility sensor (closed-loop stimulation) improve rate-adaptive pacing and resolve symptoms during high-intensity exercise in a patient with AF and AV node ablation?
Case Report (n=1)
Does an impedance-derived contractility sensor (closed-loop stimulation) improve rate-adaptive pacing and resolve symptoms during high-intensity exercise in a patient with AF and AV node ablation?
Answers are grounded in this paper and cite their sources.
This case report provides the first description of the utility of closed-loop stimulation for effective rate-adaptive pacing during high-intensity exercise in a patient with AF.
Effective rate-adaptive pacing may be difficult in the presence of atrial fibrillation (AF), and is important during high-intensity exercise. This case presents a 74-year-old elite cyclist with AF and a biventricular pacemaker after atrioventricular (AV) node ablation. He reported sudden breathlessness due to heart rate drops, caused by breaching the artefact threshold on the minute-ventilation sensor. He was exchanged to a generator with an impedance-derived contractility sensor (closed-loop stimulation), resulting in resolution of symptoms, and no further rate drops. This is the first description of the utility of closed-loop stimulation in high-intensity exercise.
No takes yet. Share an insight, caveat, or question.
Procter et al. (2023) conducted a case report in Atrial fibrillation and biventricular pacemaker after AV node ablation (n=1). Generator with an impedance-derived contractility sensor (closed-loop stimulation) vs. Minute-ventilation sensor (previous generator) was evaluated on Resolution of symptoms and rate drops. Exchanging a pacemaker generator to one with an impedance-derived contractility sensor (closed-loop stimulation) resolved sudden breathlessness and heart rate drops in a 74-year-old elite cyclist.