Key result
Leadless ECG matches surface ECG for dual-chamber ICD pacing threshold measurements.
Why the study?
A surface ECG for pacing threshold measurements during routine ICD follow-up can be cumbersome, motivating evaluation of an intracardiac electrogram alternative.
Does the use of an intrathoracic far-field electrogram (LLECG) provide comparable pacing threshold measurements to surface ECG in patients with dual chamber ICDs?
Observational (n=888)
Does the use of an intrathoracic far-field electrogram (LLECG) provide comparable pacing threshold measurements to surface ECG in patients with dual chamber ICDs?
The use of an intrathoracic far-field electrogram (LLECG) provides reliable pacing threshold measurements comparable to surface ECG, saving time during routine ICD follow-up.
Answers are grounded in this paper and cite their sources.
LLECG may simplify threshold testing in dual-chamber ICD follow-up; leaves open accuracy versus surface ECG pending larger validation studies.
BACKGROUND: A surface electrocardiogram (SECG) for pacing threshold measurements during routine implantable cardioverter-defibrillator (ICD) follow-up can be cumbersome. This study evaluated the use of an intrathoracic far-field electrogram (EGM) derived between the Can and superior vena cava (SVC) electrode -- the Leadless electrocardiogram (LLECG), in dual chamber ICDs in performing pacing threshold tests. METHODS: The LLECG was evaluated prospectively during atrial and ventricular pacing threshold testing as a substudy of the Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators trial (EMPIRIC) in which dual chamber ICDs were implanted in 888 patients. Threshold tests were conducted at 1 volt by decrementing the pulse width. Follow-up at three months compared pacing thresholds measured using LLECG with those using Lead I of the surface ECG (SECG). The timesaving afforded by LLECG was assessed by a questionnaire. RESULTS: The median threshold difference between LLECG and SECG measurements for both atrial (0.00 ms, P = 0.90) and ventricular (0.00 ms, P = 0.34) threshold tests were not significant. Ninety percent of atrial and ventricular threshold differences were bounded by +/- 0.10 ms and -0.10 to +0.04 ms, respectively. We found that 99% of atrial and ventricular thresholds tests at six and 12 months attempted using LLECG were successfully completed. The questionnaire indicated that 65% of healthcare professionals found LLECG to afford at least some timesaving during device follow-ups. CONCLUSION: Routine follow-up can be performed reliably and expeditiously in dual chamber Medtronic (Minneapolis, MN, USA) ICDs using LLECG alone, resulting in overall timesaving.
No takes yet. Share an insight, caveat, or question.
Michael et al. (2007) conducted an observational in Patients with dual chamber implantable cardioverter-defibrillators (ICDs) (n=888). Leadless electrocardiogram (LLECG) vs. Lead I of the surface ECG (SECG) was evaluated on Pacing threshold difference between LLECG and SECG measurements (Median difference 0.00 ms, p=0.90 (atrial), 0.34 (ventricular)). Leadless electrocardiogram (LLECG) for pacing threshold tests in dual chamber ICDs showed no significant difference compared to surface ECG.