We report a significant improvement of spectral properties of a cadmium telluride (CdTe) detector. This was accomplished via the use of a high quality CdTe crystal, where high Schottky barrier for the holes on a CdTe surface was formed by using a low work-function metal, indium. With a 2/spl times/2 mm/sup 2/ detector at a thickness of 0.5 mm: the leakage current is measured to be 0.7 nA at room temperature (20/spl deg/C) and below 1 pA at -70/spl deg/C for 400 V bias voltage. The low leakage current allows us to operate the detector at a higher bias voltage than for previous CdTe detectors. The energy resolution we achieved at room temperature is 1.1-2.5 keV FWHM from the energy range of 2 keV to 150 keV at 20/spl deg/C without any charge-loss correction electronics. At -70/spl deg/C, we obtained an energy resolution of 1.0 keV FWHM at 122 keV and 2.1 keV FWHM at 662 keV.
No takes yet. Share an insight, caveat, or question.
Matsumoto et al. (1998) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: