Pixel radiation detectors made on epitaxial silicon carbide are presented. Two small-format prototypes have been fabricated: a 4times4 matrix with pixels of 400times400 mum2and 6times6 matrix with 200times200 mum2pixels. Typical leakage currents between 3 fA and 10 fA for the larger pixel size, and between 0.1 fA and 4 fA for the small pixels have been measured at 27 degC, corresponding to current densities between 0.25 and 10 pA/cm2. In terms of equivalent noise charge, the contribution of most of the pixels is lower than 1 electron root mean square (rms) up to peaking times of tens of mus of pulse shaping. These pixel detectors are ready for applications in ultimate-resolution X-ray spectroscopic imaging at room temperature when a suitable ultra low noise front-end electronics, presently not available, will be developed. An analysis of the experimental data on these detectors coupled to front-end transistors of commercial CMOS technologies is presented, indicating that a noise level of around 4-electrons rms can be achieved at room temperature, limited by the front-end 1/f noise. The conditions to fully exploit the SiC pixel capabilities are quantitatively analysed by considering CMOS technologies for front-end and preamplifier design and the continuous progress in the SiC growing processes
No takes yet. Share an insight, caveat, or question.
Bertuccio et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: