Key result
Continuous-flow VAD transit reduces ultrasound contrast signal intensity by ~22% compared with prepump levels.
Why the study?
Contrast agents improve transthoracic echocardiography image quality in patients with ventricular assist devices, but their hydrodynamic fragility raises concerns about microsphere destruction during passage through continuous flow VADs.
Does passage through a continuous flow ventricular assist device reduce contrast microsphere concentration in an in vitro model?
Does passage through a continuous flow ventricular assist device reduce contrast microsphere concentration in an in vitro model?
Contrast microspheres undergo significant destruction when passing through a continuous flow ventricular assist device, which may impact the efficacy of contrast-enhanced echocardiography in these patients.
Answers are grounded in this paper and cite their sources.
Contrast may retain integrity after VAD passage in bench models; leaves open whether it improves TTE imaging in supported patients.
Objectives. Transthoracic echocardiography (TTE) is fundamental in managing patients supported with ventricular assist devices (VAD). However imaging can be difficult in these patients. Contrast improves image quality but they are hydrodynamically fragile agents. The aim was to assess contrast concentration following passage through a VAD utilising a mock circulation loop (MCL). Methods. Heartware continuous flow (CF) VAD was incorporated into a MCL. Definity® contrast was infused into the MCL with imaging before and after CF-VAD. 5 mm2 regions of interest were used to obtain signal intensity (decibels), as a surrogate of contrast concentration. Results. Four pump speeds revealed significant reduction in contrast signal intensity after CF-VAD compared to before CF-VAD (all p<0.0001 ). Combined pre- and postpump data at all speeds showed a 22.2% absolute reduction in contrast signal intensity across the CF-VAD (14.8 ± 0.8 dB prepump versus 11.6 ± 1.4 dB postpump; p<0.0001 ). Mean signal intensity reduction at each speed showed an inverse relationship between speed and relative reduction in signal intensity. Conclusion. Contrast microsphere transit through a CF-VAD within a MCL resulted in significant reduction in signal intensity, consistent with destruction within the pump. This was evident at all CF-VAD pump speeds but relative signal drop was inversely proportional to pump speed.
No takes yet. Share an insight, caveat, or question.
Platts et al. (2017) studied Ventricular assist device imaging. Passage through a continuous flow ventricular assist device (CF-VAD) vs. Before CF-VAD (prepump) was evaluated on Contrast signal intensity (decibels) (22.2% absolute reduction, p=<0.0001). Transit of contrast microspheres through a continuous flow ventricular assist device reduced signal intensity.