Directly collected multiple breath washout data in corrected 3.3.1 software differed significantly from migrated and rerun legacy data, with 37% of participants showing a relative difference >10%.
Why the study?
Is migrating or rerunning legacy MBWN2 data equivalent to directly collecting data in the new corrected software?
Observational (n=19)
Is migrating or rerunning legacy MBWN2 data equivalent to directly collecting data in the new corrected software?
Directly collecting MBWN2 data on corrected software yields significantly different results compared to migrating or rerunning legacy data, necessitating a new baseline for clinical interpretation.
Significant LCI2.5 differences between direct and legacy MBW methods warrant caution in interpretation; hypothesis-generating for updated reference standards.
Recently, a cross‐talk error with commercial multiple breath nitrogen washout (MBWN2) software was discovered, which produced an absolute over‐reading of N2 of approximately 1%, i.e., 2% N2 read as 3%. This caused an extended tail to the washout, and over‐estimated lung clearance index (LCI2.5) values. Subsequently an updated and corrected software version has been released. Within the field there have been discussions on how to correct legacy data, whether to migrate or completely “rerun” raw data A‐files from the old software into the new corrected software. To our knowledge, no research has been published assessing whether either method is equivalent to directly collecting data in the new corrected software. We prospectively recruited 19 participants, 10 adult healthy controls and 9 people with cystic fibrosis (CF). MBWN2 was performed using the Exhalyzer® D first on the old 3.1.6 software and next, directly on corrected 3.3.1 software. Multiple breath washout (MBW) data directly collected in 3.3.1 was significantly different from both migrated and rerun data. A total of 7 of the 19 participants (37%; 4 CF) had a relative difference in LCI2.5 > 10% for both migrated and rerun data compared to 3.3.1 collected data. Our findings have implications for the Global Lung Initiative MBW project, which is accepting a combination of directly collected, A‐file reruns and migrated data to establish normative values. Further, caution must be used in clinical practice when comparing corrected legacy data versus 3.3.1 collected data for clinical interpretation. We recommend that a new baseline is collected directly on 3.3.1. before clinical interpretation and decisions are determined when comparing consecutive MBW tests.
No takes yet. Share an insight, caveat, or question.
Short et al. (2022) conducted an observational in Cystic fibrosis and healthy controls (n=19). Direct data collection in corrected 3.3.1 software vs. Migrated or rerun data from old 3.1.6 software was evaluated on Relative difference in LCI2.5 > 10% compared to 3.3.1 collected data. Directly collected multiple breath washout data in corrected 3.3.1 software differed significantly from migrated and rerun legacy data, with 37% of participants showing a relative difference >10%.