Key result
Adding a septal electrode cuts atrial defibrillation thresholds by ~32% versus standard configuration.
Why the study?
The standard atrial defibrillation lead configuration uses electrodes in the right atrial appendage and coronary sinus, but the effect of adding an atrial septal electrode on atrial defibrillation threshold was unknown.
Answers are grounded in this paper and cite their sources.
BACKGROUND: The standard lead configuration for internal atrial defibrillation consists of a shock between electrodes in the right atrial appendage (RAA) and coronary sinus (CS). We tested the hypothesis that the atrial defibrillation threshold (ADFT) of this RAA-->CS configuration would be lowered with use of an additional electrode at the atrial septum (SP). METHODS AND RESULTS: Sustained atrial fibrillation was induced in 8 closed-chest sheep with burst pacing and continuous pericardial infusion of acetyl-ss-methylcholine. Defibrillation electrodes were situated in the RAA, CS, pulmonary artery (PA), low right atrium (LRA), and across the SP. ADFTs of RAA-->CS and 4 other lead configurations were determined in random order by use of a multiple-reversal protocol. Biphasic waveforms of 3/1-ms duration were used for all single and sequential shocks. The ADFT delivered energies for the single-shock configurations were 1.27+/-0.67 J for RAA-->CS and 0. 86+/-0.59 J for RAA+CS-->SP; the ADFTs for the sequential-shock configurations were 0.39+/-0.18 J for RAA-->SP/CS-->SP, 1.16+/-0.72 J for CS-->SP/RAA-->SP, and 0.68+/-0.46 J for RAA-->CS/LRA-->PA. Except for CS-->SP/RAA-->SP versus RAA-->CS and RAA-->CS/LRA-->PA versus RAA+CS-->SP, the ADFT delivered energies of all of the configurations were significantly different from each other (P:<0. 05). CONCLUSIONS: The ADFT of the standard RAA-->CS configuration is markedly reduced with an additional electrode at the atrial SP.
No takes yet. Share an insight, caveat, or question.
Zheng et al. (2000) studied Sustained atrial fibrillation (n=8). Additional electrode at the atrial septum (SP) vs. Standard RAA-->CS configuration was evaluated on Atrial defibrillation threshold (ADFT) delivered energy (p=<0.05). An additional electrode at the atrial septum reduced the atrial defibrillation threshold compared to the standard RAA-->CS configuration.
Synapse has enriched 10 related papers. Consider them for comparative context: