Key result
Fourier interpolation improves MRI coronary area measurement precision ~6.5-fold, enabling detection of subtle cross-sectional changes.
Why the study?
MRI can noninvasively measure coronary endothelial function by quantifying coronary artery cross-sectional area changes, but limited spatial resolution challenges measurement precision and accuracy.
Answers are grounded in this paper and cite their sources.
Enables detection of 4-5% coronary area changes with current spiral MRI; leaves open clinical utility for endothelial function assessment pending validation.
PURPOSE: Coronary endothelial function (CEF) reflects vascular health and conventional invasive CEF measures predict cardiovascular events. MRI can now noninvasively measure CEF by quantifying coronary artery cross-sectional area changes in response to isometric handgrip exercise, an endothelial-dependent stressor. Area changes (10 to 20% in healthy; 2 to -12% in impaired vessels) are only a few imaging voxels because of MRI's limited spatial resolution. Here, with numerical simulations and phantom studies, we test whether Fourier interpolation enables sub-pixel area measurement precision and determine the smallest detectable area change using spiral MRI. METHODS: In vivo coronary SNR with the currently used CEF protocol at 3T was measured in 7 subjects for subsequent in vitro work. Area measurements of circular vessels were simulated by varying partial volume, vessel diameter, voxel size, SNR, and Fourier interpolation factor. A phantom with precision-drilled holes (diameters 3-3.42 mm) was imaged 10 times with the current CEF protocol (voxel size, Δx = 0.89 mm) and a high-resolution protocol (Δx = 0.6 mm) to determine precision, accuracy, and the smallest detectable area changes. RESULTS: In vivo coronary SNR ranged from 30-76. Eight-fold Fourier interpolation improved area measurement precision by a factor 6.5 and 4.9 in the simulations and phantom scans, respectively. The current CEF protocol can detect mean area changes of 4-5% for SNR above 30, and 3-3.5% for SNR above 40 with a higher-resolution protocol. CONCLUSION: Current CEF spiral MRI with in vivo SNR allows detection of a 4-5% area change and Fourier interpolation improves precision several-fold to sub-voxel dimensions.
No takes yet. Share an insight, caveat, or question.
Schär et al. (2018) studied Coronary endothelial function assessment (n=7). Fourier interpolation for spiral MRI vs. Images without Fourier interpolation was evaluated on Improvement in area measurement precision and smallest detectable area change (limit of detection). Eight-fold Fourier interpolation improved area measurement precision by a factor of 6.5 in simulations and enabled current spiral MRI protocols to detect coronary cross-sectional area changes of 4% to 5% at in vivo SNR levels.
Synapse has enriched 10 related papers. Consider them for comparative context: