Key result
Lung ultrasound tracks fluid index and predicts heart failure deterioration with ~91% PPV.
Why the study?
Both ultrasound lung comets (ULCs) and implantable defibrillator-measured intrathoracic impedance assess pulmonary congestion in HF patients. This study evaluated the agreement between defibrillator-detected impedance decreases and the presence of ULCs, and compared their performance in predicting HF worsening.
Does chest ultrasound assessment of ultrasound lung comets correlate with defibrillator-measured intrathoracic impedance and predict clinical heart failure deterioration in heart failure patients?
Cohort (n=23)
Does chest ultrasound assessment of ultrasound lung comets correlate with defibrillator-measured intrathoracic impedance and predict clinical heart failure deterioration in heart failure patients?
Chest ultrasound detection of lung comets strongly correlates with implantable defibrillator intrathoracic impedance and provides high positive predictive value (91%) for identifying clinical heart failure decompensation.
Answers are grounded in this paper and cite their sources.
May aid noninvasive HF monitoring in device patients; leaves open need for larger prospective validation before routine adoption.
PURPOSE: Echographic examination of the lung surface may reveal multiple ultrasound lung comets (ULCs) originating from water-thickened interlobular septa. These images were demonstrated to be useful for noninvasive assessment of interstitial pulmonary edema. Similarly, the correlation between implantable defibrillator-measured intrathoracic impedance and pulmonary capillary wedge pressure (PCWP) was demonstrated in heart failure (HF) patients. The aims of this analysis were to assess the agreement between defibrillator-detected impedance decrease and the presence of ULCs, as well as to compare the performance of the impedance-detection algorithm and the ULCs assessment in predicting HF worsening. METHODS AND RESULTS: We studied 23 HF patients implanted with a defibrillator capable of intrathoracic impedance measurement and alerting for fluid accumulation diagnosis. At regular follow-up and at visits for HF decompensation or device alert, clinical status was assessed, chest ultrasound was performed, and PCWP was noninvasively estimated with Doppler echocardiography. During 23 ± 11 months of follow-up, 45 paired assessments of defibrillator-measured intrathoracic impedance and ULCs were performed. The number of ULCs resulted significantly correlated to the paired PCWP estimations (r = 0.917, P < 0.001) and to the impedance fluid index (r = 0.669, P < 0.001). During follow-up, 12 episodes of HF decompensation were reported in 12 patients. The impedance alert detected clinical HF deterioration with 92% sensitivity and 69% positive predictive value. The presence of ≥5 ULCs showed 83% sensitivity and 91% positive predictive value. CONCLUSIONS: These data demonstrate the correlation between intrathoracic impedance and the number of ULCs at chest ultrasound, and a good agreement between the defibrillator warning for fluid index and the detection of multiple ULCs.
No takes yet. Share an insight, caveat, or question.
Maines et al. (2011) conducted a cohort in Heart failure (n=23). Ultrasound lung comets (ULCs) assessment vs. Defibrillator-measured intrathoracic impedance alert was evaluated on Prediction of clinical heart failure deterioration / correlation between ULCs and impedance fluid index (r = 0.669, p=<0.001). Ultrasound lung comets correlated with the impedance fluid index and detected HF deterioration with 83% sensitivity and 91% PPV vs 92% sensitivity and 69% PPV for impedance alerts.