Key result
Defibrillator shocks for atrial arrhythmias in patients with a Ventak P2 pulse generator and Endotak C lead caused transient marked widening of intracardiac ventricular complexes resembling VT.
Case Report (n=3)
Clinicians should be aware that defibrillator shocks for atrial arrhythmias can cause transient distortion of intracardiac electrograms that mimic ventricular tachycardia.
Answers are grounded in this paper and cite their sources.
Supports awareness of shock-induced electrogram distortion mimicking VT; leaves open generalizability to contemporary devices.
In three patients with a defibrillator system consisting of a Ventak P2 pulse generator and an Endotak C transvenous lead, we observed distortion of intracardiac electrograms following defibrillator shocks for atrial arrhythmias. There was a transient marked widening of the intracardiac ventricular complexes resembling ventricular tachycardia. This phenomenon should be recognized when evaluating arrhythmic episodes.
No takes yet. Share an insight, caveat, or question.
Khalighi et al. (1997) conducted a case report in Atrial tachyarrhythmias (n=3). Defibrillator shocks was evaluated on Distortion of intracardiac electrograms. Defibrillator shocks for atrial arrhythmias in patients with a Ventak P2 pulse generator and Endotak C lead caused transient marked widening of intracardiac ventricular complexes resembling VT.