Answers are grounded in this paper and cite their sources.
The effect of guidewire bias on angled-lesion ablation by rotational atherectomy (RA) was assessed by measuring the changes in vertical lumen diameter, horizontal lumen diameter and the intima-media thickness of the coronary artery, using intravascular ultrasound in 10 lesions with an angle greater than 10 degrees. The vertical and horizontal diameters significantly increased after RA. The intima-media thickness at the 4 orthogonal sites significantly decreased. There was a significant positive correlation between vertical diameter change and angle (r=0.642, p=0.045), but none between horizontal diameter change and angle. There was no correlation between intima-media thickness change at 0 degrees and angle; however, at 180 degrees there was a tendency to correlation with angle (r=0.602, p=0.066). These data suggest that in cases of angled lesions, the increase in vertical lumen diameter is caused more by ablation of the 180 degrees wall than by that of the 0 degrees wall, which is brought about by guidewire bias toward the vascular wall at 180 degrees.
No takes yet. Share an insight, caveat, or question.
Oishi et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: