Key result
Automated atrial threshold testing triggers ventricular fibrillation by inducing temporary RV-only pacing in CRT-D.
Case Report (n=1)
Highlights a critical pro-arrhythmic risk of automated atrial threshold testing causing RV-only pacing in susceptible CRT patients.
Answers are grounded in this paper and cite their sources.
May warrant caution with automated atrial threshold testing in CRT-D patients; leaves open incidence and mitigation in broader populations.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Jama et al. (2022) conducted a case report in Cardiac sarcoidosis with severe LV systolic dysfunction and CRT-D (n=1). Automated atrial lead threshold testing (Capture Management) vs. Biventricular pacing was evaluated on Ventricular fibrillation or tachycardia requiring shock therapy. Automated atrial threshold testing caused temporary RV-only pacing, triggering ventricular fibrillation in a susceptible patient with a CRT-D.