Key result
The Panna wire improved the operative success rate of percutaneous atrial septal defect closure by junior doctors compared to a conventional guidewire.
Why the study?
Echo-guided percutaneous procedures are reliable, but the learning curve is difficult for junior doctors, prompting evaluation of a novel guidewire (Panna wire) for percutaneous ASD closure under transthoracic echocardiography guidance only.
Does the Panna wire improve operative success rate compared to a conventional guidewire in patients undergoing percutaneous atrial septal defect closure by junior doctors?
RCT (n=100)
randomized
Yes
Does the Panna wire improve operative success rate compared to a conventional guidewire in patients undergoing percutaneous atrial septal defect closure by junior doctors?
The novel Panna guidewire significantly improves procedural success and efficiency for junior doctors performing nonfluoroscopic, echo-guided atrial septal defect closure.
Answers are grounded in this paper and cite their sources.
May support Panna wire use; hypothesis-generating and requires randomized confirmation before practice change.
Background: Echo-guided percutaneous procedures have been reported reliable and advantageous. However, the learning curve is difficult for junior doctors. We aimed to evaluate the safety and efficacy of a novel guidewire (Panna wire) in percutaneous atrial septal defect closure under transthoracic echocardiography guidance only. Methods: The Panna wire is designed for echo-guide procedure with a retractable spindle-shaped tip. A multicenter, randomized, controlled trial was conducted to evaluate the safety and efficacy of the Panna wire versus the conventional guidewire for junior doctors with <100 cases experience. The primary outcome was operative success rate. The secondary outcomes were incidence of major adverse events, operation time, time needed to enter the left atrium, number of arrhythmia episodes, number of misguidance to tricuspid valve, and incidence of peripheral vascular complications. Results: Between July 2018 and September 2019, 100 patients with atrial septal defect were randomized to either the Panna wire group (n=52) or the conventional wire group (n=48) at 3 centers. The baseline clinical characteristics were similarly distributed. The operative success rate (primary outcome) was 100% in the Panna wire group versus 68.75% in the conventional wire group ( P <0.001). No major adverse events occurred in either group. Significant differences in favor of the Panna wire group were found in operation time ( P =0.004), time needed to enter the left atrium ( P <0.001), number of arrhythmia episodes ( P <0.001), and number of misguidance to tricuspid valve ( P =0.005). Conclusions: The Panna wire is safe and effective and reduces the learning curve in percutaneous atrial septal defect closure under transthoracic echocardiography guidance only. Registration: URL: https://www.clinicaltrials.gov . Unique identifier: NCT04096924.
No takes yet. Share an insight, caveat, or question.
Kong et al. (2020) conducted an RCT in atrial septal defect (n=100). Panna wire vs. conventional guidewire was evaluated on operative success rate (p=<0.001). The Panna wire improved the operative success rate of percutaneous atrial septal defect closure by junior doctors compared to a conventional guidewire.
Synapse has enriched 10 related papers. Consider them for comparative context: