Key result
Radiofrequency ablation using a contact force-sensing catheter in a low-volume center resulted in 0 complications related to catheter manipulation or energy delivery among 39 patients.
Why the study?
Does percutaneous radiofrequency ablation using a contact force-sensing catheter ensure procedural safety in patients with atrial fibrillation treated at a low-volume centre?
Observational (n=39)
No
Does percutaneous radiofrequency ablation using a contact force-sensing catheter ensure procedural safety in patients with atrial fibrillation treated at a low-volume centre?
The use of a contact force-sensing catheter during atrial fibrillation ablation in a low-volume center appears safe, helping to avoid excessive contact forces and procedural complications.
Answers are grounded in this paper and cite their sources.
May support safe use of contact force-sensing catheters in low-volume centers; hypothesis-generating and requires larger prospective validation.
The tip-to-tissue contact force (CF) has been identified as a potential determinant of lesion quality during radiofrequency (RF) ablation. The aim of this paper is to report the experience of a single low-volume centre in the atrial fibrillation (AF) ablation procedure with an RF catheter capable of measuring this parameter. CF data and their possible implications on patient safety are presented. Thirty-nine consecutive patients suffering of paroxysmal or permanent AF received percutaneous ablation with the novel catheter studied. Procedural characteristics, CF applied and safety events related to the procedure were reported. During RF application the mean CF value was 17 ± 3 g, with a maximum mean value of 37 ± 8 g. CF value never exceeded 62 g and in the 74% of the RF applications ranged between 10 g and 30 g. No complication related to the catheter manipulation or to the energy delivered was observed. This study of a single centre with a low level of experience in AF ablation suggests that the ability to measure CF may provide additional useful information to the operator. It ensures uniform ablations, with little variability in the catheter manipulations, and it avoids excessive contact forces increasing the patient safety.
No takes yet. Share an insight, caveat, or question.
Vaccari et al. (2016) conducted an observational in Atrial fibrillation (n=39). Radiofrequency ablation with a contact force-sensing catheter was evaluated on Complications related to catheter manipulation or energy delivered. Radiofrequency ablation using a contact force-sensing catheter in a low-volume center resulted in 0 complications related to catheter manipulation or energy delivery among 39 patients.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: