Key result
The TM-2655 oscillometric device achieved British Hypertension Society grade A/A, with an average difference of -1.0±5.2 mmHg for systolic and -0.9±4.7 mmHg for diastolic blood pressure.
Why the study?
Does the TM-2655 oscillometric device accurately measure blood pressure compared to a mercury sphygmomanometer in an adult population?
Observational (n=85)
Does the TM-2655 oscillometric device accurately measure blood pressure compared to a mercury sphygmomanometer in an adult population?
The TM-2655 oscillometric device meets the highest accuracy standards (grade A/A) of the British Hypertension Society protocol and is recommended for adult blood pressure measurement.
Answers are grounded in this paper and cite their sources.
Supports TM-2655 accuracy in adults; leaves open generalizability and practice change pending larger validation studies.
OBJECTIVE: To perform validation for an arm-type oscillometric TM-2655 device (A&D Company Ltd, Tokyo, Japan) for blood pressure measurement according to the British Hypertension Society protocol. METHODS: Eighty-five study participants (33 men and 52 women) were included in the study. Mean age was 52.9+/-15.0 years, systolic blood pressure range was 84-208 mmHg and diastolic blood pressure range was 48-120 mmHg. For each participant, three readings of TM-2655 were compared with sequential auscultatory measurements by two trained independent observers. The observers used a calibrated mercury sphygmomanometer and dual stethoscope. The results were graded according to the British Hypertension Society protocol 1993. RESULTS: The average difference between mercury sphygmomanometer and TM-2655 readings for systolic blood pressure was -1.0+5.2 mmHg (mean+/-SD) and for diastolic blood pressure -0.9+/-4.7 mmHg. The proportions of values agreeing to within 5, 10 and 15 mmHg were 72.5, 93.7 and 99.6% for systolic blood pressure and 78.8, 96.9 and 100% for diastolic blood pressure between the observers and the device (A/A British Hypertension Society grade). CONCLUSIONS: The TM-2655 device achieved British Hypertension Society grade A/A and therefore can be recommended for blood pressure measurement in an adult population.
No takes yet. Share an insight, caveat, or question.
Kobalava et al. (2006) conducted an observational in Blood pressure measurement (n=85). TM-2655 oscillometric device vs. Mercury sphygmomanometer was evaluated on Average difference in systolic and diastolic blood pressure readings (MD -1.0 mmHg (SBP) and -0.9 mmHg (DBP)). The TM-2655 oscillometric device achieved British Hypertension Society grade A/A, with an average difference of -1.0±5.2 mmHg for systolic and -0.9±4.7 mmHg for diastolic blood pressure.