OBJECTIVE: To assess the Dixtal DX2710 automated oscillometric device used for blood pressure measurement according to the protocols of the BHS and the AAMI. METHODS: Three blood pressure measurements were taken in 94 patients (53 females 15 to 80 years). The measurements were taken randomly by 2 observers trained to measure blood pressure with a mercury column device connected with an automated device. The device was classified according to the protocols of the BHS and AAMI. RESULT: The mean of blood pressure levels obtained by the observers was 148+/-38/93+/-25 mmHg and that obtained with the device was 148+/-37/89+/-26 mmHg. Considering the differences between the measurements obtained by the observer and those obtained with the automated device according to the criteria of the BHS, the following classification was adopted: "A" for systolic pressure (69% of the differences < 5; 90% < 10; and 97% < 15 mmHg); and "B" for diastolic pressure (63% of the differences < 5; 83% < 10; and 93% < 15 mmHg). The mean and standard deviation of the differences were 0+/-6.27 mmHg for systolic pressure and 3.82+/-6.21 mmHg for diastolic pressure. CONCLUSION: The Dixtal DX2710 device was approved according to the international recommendations.
Building similarity graph...
Analyzing shared references across papers
Loading...
Mano et al. (Sun,) studied this question.
synapsesocial.com/papers/6a1bc29300ee29383e9cd6d1 — DOI: https://doi.org/10.1590/s0066-782x2002001500006
Gisele Machado Peixoto Mano
Vodafone (Portugal)
Viviane Fernandes de Souza
Universidade Brasil
Ângela M. G. Pierin
Universidade de São Paulo
Arquivos Brasileiros de Cardiologia
Universidade de São Paulo
Fundação Faculdade de Medicina
Microsoft (Brazil)
Building similarity graph...
Analyzing shared references across papers
Loading...