Does a mechanical heart valve prosthesis with a tilting disc cause hemolysis in an in vitro model?
Does a mechanical heart valve prosthesis with a tilting disc cause hemolysis in an in vitro model?
In vitro hemolysis testing using a lysis number can estimate erythrocyte survival time in patients with mechanical heart valves, showing agreement with clinical observations.
Answers are grounded in this paper and cite their sources.
Supports in vitro lysis testing to estimate hemolysis with tilting-disc valves; leaves open validation in modern prostheses and patients.
Hemolytic and subhemolytic blood damage by mechanical heart valve prostheses have been observed in both clinical and in vitro investigations. A direct comparison between these studies is not possible. Nevertheless the transfer of some in vitro results to the behaviour of the valve in situ may be performed considering the similarity principle. This requires the use of dimensionless similarity numbers such as the plasma's hemoglobin concentration (PHb) or others, instead of dimensioned parameters. To evaluate the in vitro hemolysis of valve prosthesis a test chamber filled with human banked blood was used. An artificial ventricle ensuring an oscillatory flow through the valve was also used. The rise of PHb was evaluated in terms of a similarity number, called the lysis number. This number describes the probability of destroying a single red blood cell participating once in the hemolytic process under consideration. The lysis number, a Björk-Shiley valve (TAD 29), was found to be in the order of 2 x 10(-4). From this, the survival time of erythrocytes in patients with an artificial heart valve was estimated. It was found to be in the order of 20 d of T50 Cr in agreement with clinical results.
No takes yet. Share an insight, caveat, or question.
Wendt et al. (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: