Why the study?
How do evoked QT interval sensing pacemakers compare to vibration sensing pacemakers in rate adaptation during exercise?
How do evoked QT interval sensing pacemakers compare to vibration sensing pacemakers in rate adaptation during exercise?
Both evoked QT and vibration-sensing pacemakers exhibit specific deficiencies in physiological rate adaptation during exercise, highlighting the need for further technological development.
Answers are grounded in this paper and cite their sources.
Rate-adaptive pacemakers may underperform during exercise; leaves open the need for improved sensor algorithms before wider adoption.
Two new generation rate adaptive pacemakers, the Rhythmyx (Vitatron) using the evoked QT interval and the Legend (Medtronic) using vibration as indicators of metabolic demand, were compared for rate adaptive characteristics during different forms of exercise. While both showed improvements over previous generation pacemakers, they still show deficiencies in some aspects of rate modulation. Rhythmyx was slow to respond to changes in metabolic need and showed an "over-shoot" with increasing pacing rate on cessation of exercise. Legend was quick to adapt rate at beginning and end of exercise but showed a plateaus of rate modulation during the period of slowly increasing workload. Legend also showed only modest rate adaptation to changes in treadmill gradient and to bicycle ergometer exercise. Further developments in pacemaker technology are required if physiological rate adaptation to exercise is required.
No takes yet. Share an insight, caveat, or question.
Bellamy et al. (1992) studied this question.