Why the study?
Does pulsatile flow affect the radial compliance of porcine coronary arteries ex vivo compared to steady flow?
Does pulsatile flow affect the radial compliance of porcine coronary arteries ex vivo compared to steady flow?
Radial compliance of coronary arteries is significantly reduced under pulsatile flow compared to steady flow, which has important implications for understanding stent biomechanics.
Answers are grounded in this paper and cite their sources.
Pulsatile flow data should inform but not guide stent-artery modeling; leaves open translation to human in vivo coronary biomechanics.
Besides visual assessment of the intra-vascular appearance, intravascular optical coherence tomography (IVOCT) enables precise diameter measurements within the coronary arteries. The current study presents an upgraded test setup allowing the investigation of the elastic properties of porcine coronary arteries ex vivo under pulsatile flow conditions. IV-OCT imaging was performed within the left circumflex artery of a porcine heart applying a normotensive and a hypertensive pressure regime at different pulse rates. Radial compliance was derived from the luminal diameter as well as intra-arterial pressure measurements. Test results show a significant reduction of radial compliance under pulsatile conditions, when compared to reference measurements under steady flow. The mean radial compliance decreased with increasing pulse rate, which can be attributed to the viscoelastic properties of the arterial wall.
No takes yet. Share an insight, caveat, or question.
Brandt-Wunderlich et al. (2022) studied this question.