Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension protocol

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does the UA-787 oscillometric blood pressure monitor accurately measure blood pressure compared to the mercury standard in human subjects?

Population

33 subjects

Comparison

UA-787 oscillometric blood pressure monitor vs Mercury standard (observers)

Design

Other

Authors

DLDaniele LongoUniversity of VeronaGTGianluca ToffaninPsychiatric Medicine AssociatesRGRaffaella GarbelottoOspedale Civile di Vittorio Veneto

Discussion

Loading...

Member takes

Implication

May support UA-787 accuracy in practice; leaves open confirmation in larger, diverse populations.

Structured PICO

Does the UA-787 oscillometric blood pressure monitor accurately measure blood pressure compared to the mercury standard in human subjects?

P
Population
33 subjects
I
Intervention
UA-787 oscillometric blood pressure monitor
C
Comparator
Mercury standard (observers)
O
Outcome
Accuracy of blood pressure measurement (difference between device and observers)surrogate

The UA-787 oscillometric blood pressure monitor meets the European Society of Hypertension accuracy requirements, validating its clinical use.

Cite This Study

Longo et al. (2003) studied this question.

synapsesocial.com/papers/6a87a8a7b438de75f6975258https://doi.org/10.1097/00126097-200304000-00006

Topics

Hypertension management
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1THE ACCURACY AND PRECISION OF THE ACCUTRACKER AMBULATORY BLOOD PRESSURE MONITOR1990 · 33 citations
  2. 2Arterial Stiffness as Underlying Mechanism of Disagreement Between an Oscillometric Blood Pressure Monitor and a Sphygmomanometer2000 · 173 citations
  3. 3Determinants of the limits of agreement between the sphygmomanometer and the SpaceLabs 90207 device for blood pressure measurement in healthy volunteers and insulin-dependent diabetic patients1998 · 53 citations
  4. 4Assessment of repeatability and correlates of arterial compliance2002 · 38 citations
  5. 5Comparison of agreement between different measures of blood pressure in primary care and daytime ambulatory blood pressure2002 · 172 citations