Key result
A smartphone-based wireless blood pressure monitoring prototype demonstrated >97% accuracy compared to standard devices, with mean differences of less than 5 mmHg.
Why the study?
Does a smartphone-based wireless blood pressure monitoring system accurately measure blood pressure compared to standard sphygmomanometers in healthy subjects?
Does a smartphone-based wireless blood pressure monitoring system accurately measure blood pressure compared to standard sphygmomanometers in healthy subjects?
A newly developed Bluetooth-enabled smartphone blood pressure monitoring system meets AAMI accuracy criteria when compared to standard devices.
Answers are grounded in this paper and cite their sources.
Warrants caution before clinical use of smartphone BP prototypes; leaves open their role in hypertension care pending larger validation studies.
In medical practice, blood pressure is intensively studied parameter and often measured for diagnosis at primary healthcare level. In this paper, the state-of-the-art Bluetooth technology has been used for designing an innovative wireless blood pressure monitoring system to display the systolic and diastolic blood pressure values on Android-based smartphone. A pilot study was undertaken on 12 healthy subjects to validate the performance of developed prototype by comparing with OMRON HEM-7111 and mercury sphygmomanometer. Comparison shows an accuracy of 98.63% and 97.13% with OMRON HEM-7111 and 98.99% and 98% with mercury sphygmomanometer for systolic blood pressure and diastolic blood pressure, respectively. Bias value in Bland-Altman plots gives good level of agreement within 95% confidence limits. The developed prototype device mean and standard deviation of differences is less than 5 and 8 mmHg, respectively, and it fulfills the recommendation criteria as per the requirement of the Association for Advancement of Medical Instruments protocol.
No takes yet. Share an insight, caveat, or question.
Singh et al. (2016) studied Healthy subjects (n=12). Smartphone-based wireless blood pressure monitoring system vs. OMRON HEM-7111 and mercury sphygmomanometer was evaluated on Accuracy of systolic and diastolic blood pressure measurement. A smartphone-based wireless blood pressure monitoring prototype demonstrated >97% accuracy compared to standard devices, with mean differences of less than 5 mmHg.