Key result
Red cell destruction was significantly lower with a non-overlapping ball valve (0.29%) compared to an overlapping ball valve (0.73%) in an experimental test chamber.
Why the study?
Does a non-overlapping closing mechanism reduce haemolysis compared to an overlapping closing mechanism in ball valves?
Does a non-overlapping closing mechanism reduce haemolysis compared to an overlapping closing mechanism in ball valves?
Non-overlapping ball valves cause significantly less in vitro haemolysis compared to overlapping ball valves.
Answers are grounded in this paper and cite their sources.
May lessen experimental haemolysis; leaves open whether non-overlapping designs improve prosthetic valve outcomes.
Overlapping and non-overlapping closing mechanisms of the ball valve type were evaluated in regard to rates of haemolysis. the test chamber used in this experimental study permitted an almost physiological valve function in a minimal volume of human whole blood. An overlapping closing mechanism was represented by the Starr-Edwards aortic silastic ball valve and a non-overlapping one by the Smeloff-Cutter aortic silastic ball valve.Red cell destruction was significantly lower with a non-overlapping ball than with an overlapping one on valve closure. the magnitude of red cells destroyed in 80 ml of blood/hour was, on average, 0.29 % and 0.73 % respectively. the Bjork-Shiley tilting disc valve was tested as a control and accounted for a red cell destruction of 0.20%.Assuming a linear progression of haemolysis with time, a blood volume of 4 800 ml, and a normal erythrocyte survival time of 120 days, the red cell destruction attributable to the prosthetic closing mechanisms could be extrapolated to in uiuo con...
No takes yet. Share an insight, caveat, or question.
Fortune et al. (1975) studied this question. Non-overlapping closing mechanism (Smeloff-Cutter aortic silastic ball valve) vs. Overlapping closing mechanism (Starr-Edwards aortic silastic ball valve) was evaluated on Red cell destruction in 80 ml of blood/hour. Red cell destruction was significantly lower with a non-overlapping ball valve (0.29%) compared to an overlapping ball valve (0.73%) in an experimental test chamber.
Synapse has enriched 10 related papers. Consider them for comparative context: