Key result
Ultrasonic QFM measures common carotid artery flow with under ~7% error and detects physiological decreases.
Why the study?
Can an ultrasonic quantitative flow measurement system accurately measure common carotid artery flow volume noninvasively?
Observational
Can an ultrasonic quantitative flow measurement system accurately measure common carotid artery flow volume noninvasively?
The ultrasonic quantitative flow measurement system provides an accurate, noninvasive method for measuring common carotid artery flow volume, which varies with age, posture, and sleep stage.
Answers are grounded in this paper and cite their sources.
Requires prospective validation before clinical use; leaves open its role in carotid flow research.
Ultrasonic quantitative flow measurement system (QFM) was developed to measure an absolute flow volume of common carotid artery (CCA) noninvasively as one of the cerebral circulation tests.It is verified from the model experiment that the output is independent of the depth of vessel and the angle between beam and flow.Moreover, the accuracy is confirmed by the comparison with the electromagnetic flow meter; the error is less than •}7%.However, it is required for accurate and reproducible measurements to handle the probe properly to detect an adequate amplitude of wall echo and to set the gate position at the vessel inner surface echo.The following clinical results have been obtained: 1) There is no significant difference in CCA flow volume between right and left sides.2) The mean value of male/female is 1.058.3)The CCA flow volume is decreased with age: 8.98m1/s in 20s and 8.03m//s in 70s.4) Body position : The rapid standing from supine position decreases the CCA flow volume; the ratio ranged from -3.3% to -22.2%.5) It is suspected that the CCA flow is independent of the blood pressure.6) CCA flow volume in sleeping stage decreases 10% in REM and 20% in non-REM.
No takes yet. Share an insight, caveat, or question.
KODAIRA et al. (1988) conducted an observational in Healthy subjects and physiological variations. Ultrasonic quantitative flow measurement system (QFM) vs. Electromagnetic flow meter and physiological baseline states was evaluated on Accuracy of common carotid artery flow volume measurement. The ultrasonic quantitative flow measurement system measured common carotid artery absolute flow volume with an error of less than ±7% and detected physiological decreases with age, standing, and sleep.