Key result
Z-5 balloon atrial septostomy risks balloon tip detachment upon pullback across unusually thick atrial septums.
Why the study?
The safety and performance of the Z-5 atrial septostomy catheter in patients with unusually thick atrial septum were not well characterized.
Case Report (n=1)
Caution is warranted when using the Z-5 catheter for balloon atrial septostomy in patients with an unusually thick atrial septum due to the risk of the balloon tip tearing off.
Answers are grounded in this paper and cite their sources.
Supports caution with Z-5 in thick septa; leaves open optimal device selection for neonatal septostomy.
Creation of atrial communication was performed in a newborn with critical aortic stenosis. After the success of the initial creation, balloon atrial septostomy using Z‐5 catheter was performed. When catheter was pulled back, the tip of the balloon was torn off. This experience could be considered as noteworthy when using this catheter in patients with unusually thick atrial septum. Cathet Cardiovasc Intervent 2001;52:500–503.
No takes yet. Share an insight, caveat, or question.
Akagi et al. (2001) conducted a case report in Critical aortic stenosis (n=1). Balloon atrial septostomy using Z-5 catheter was evaluated on Torn-off balloon tip. Balloon atrial septostomy using a Z-5 catheter in a newborn with critical aortic stenosis resulted in the balloon tip tearing off upon pullback, highlighting risks with unusually thick atrial septums.