Key result
A double-valved biomechanical heart enhanced pulmonary blood flow by 1,235 +/- 526 mL/min compared to 113 +/- 37 mL/min with a valveless model in goats with induced heart failure.
Why the study?
Does a valved biomechanical heart improve circulatory flow compared to a valveless model in a goat model of heart failure?
Does a valved biomechanical heart improve circulatory flow compared to a valveless model in a goat model of heart failure?
The incorporation of two valves in a biomechanical heart model is essential to achieve adequate circulatory support in a failing heart model.
Does not inform clinical practice; extends animal evidence for valved biomechanical support but leaves translation open.
Experimental valveless muscular blood pumps (biomechanical hearts) in goats can pump more than 1 L.min(-1), but due to a high pendulum volume, no significant flow contribution to the circulation is gained. Thus valved and valveless biomechanical hearts were compared for efficacy. Heart failure was induced in 5 adult Bore goats by repeated intracoronary embolization. A valved and balloon-equipped pumping chamber was integrated into the descending aorta, simulating standard biomechanical circulatory support. The valveless biomechanical heart supported a failing heart with a baseline cardiac output of 2,670 +/- 710 mL.min(-1) by contributing additional flow of 113 +/- 37 mL.min(-1). The biomechanical heart model incorporating an outlet valve offered an additional 304 +/- 126 mL.min(-1), and the use of 2 valves significantly enhanced pulmonary blood flow by 1,235 +/- 526 mL.min(-1). The use of 2 valves in biomechanical hearts seems to be essential to achieve adequate circulatory support. Double-valved biomechanical hearts driven by an appropriate skeletal muscle ventricle may contribute to the therapy of heart failure.
No takes yet. Share an insight, caveat, or question.
Guldner et al. (2009) studied Heart failure (n=5). Valved biomechanical heart vs. Valveless biomechanical heart was evaluated on Additional flow contribution to the circulation. A double-valved biomechanical heart enhanced pulmonary blood flow by 1,235 +/- 526 mL/min compared to 113 +/- 37 mL/min with a valveless model in goats with induced heart failure.