Key result
Laser lead extraction is not associated with increased tricuspid insufficiency, occurring in only ~6% post-procedure.
Why the study?
Laser lead extraction has become more prevalent but its association with tricuspid insufficiency remains uncertain.
Does laser lead extraction increase the risk of tricuspid insufficiency in patients with pacemaker or ICD leads?
Observational (n=173)
No
Does laser lead extraction increase the risk of tricuspid insufficiency in patients with pacemaker or ICD leads?
Laser lead extraction does not appear to be associated with an increased risk of tricuspid insufficiency, with most cases of post-procedural TI occurring in patients with pre-existing endocarditis.
Answers are grounded in this paper and cite their sources.
May reassure on TI risk with laser extraction; leaves open need for randomized confirmation before practice change.
BACKGROUND: The use of laser lead extraction (LLE) to remove pacemaker and implantable cardiac defibrillator leads has become more prevalent in the past decade. Though the procedure is associated with a low rate of complications, LLE still poses some risks to patients. Some reports have suggested an increase in tricuspid insufficiency (TI) associated with LLE. We present a series of patients who underwent both LLE and complete evaluation for TI with echocardiographic techniques. METHODS: From August 2008 to January 2010, 173 prospective, consecutive patients underwent LLE in a single center. All patients had transesophageal echocardiograms (TEE) during the extraction. Fifty-three patients had tricuspid valve function evaluated a day before the procedure with a transthoracic echocardiogram (TTE), during the procedure with a TEE and 2 days postoperatively with a TTE. RESULTS: All 173 patients experienced no change in tricuspid valve function during the procedure with TEE. Of the 53 patients who underwent a complete TI evaluation, 38 were males (72%) and 15 females (38%), with a mean age of 69.45 ± 14.08. Mean ejection fraction was 35.82 ± 14.72. Three (6%) patients experienced TI after the procedure (two mild and one severe, all with tricuspid valve endocarditis); 16 (30%) patients were found to have TI before LLE that returned to normal valve function during or after the procedure. Thirty-four (64%) patients did not experience any significant change of the tricuspid valve performance after LLE. CONCLUSION: LLE was not associated with increased TI.
No takes yet. Share an insight, caveat, or question.
Rodriguez et al. (2013) conducted an observational in Pacemaker and implantable cardiac defibrillator leads requiring extraction (n=173). Laser lead extraction was evaluated on Tricuspid insufficiency after the procedure. Laser lead extraction was not associated with increased tricuspid insufficiency, with only 6% of evaluated patients experiencing it post-procedure.
Synapse has enriched 10 related papers. Consider them for comparative context: