Key result
Side helix active fixation LV lead extraction achieves 100% success without complications in a sheep model.
Why the study?
A cardiac LV lead with a side helix for active fixation to the coronary vein wall was developed to improve placement and reduce dislodgement rates, but its extractability had not been studied extensively.
Is the extraction of a novel left ventricular active fixation lead with a side helix safe and feasible compared to a standard unipolar LV lead in a sheep model?
Is the extraction of a novel left ventricular active fixation lead with a side helix safe and feasible compared to a standard unipolar LV lead in a sheep model?
Extraction of a novel active fixation LV lead with a side helix is safe and uncomplicated in a sheep model up to 118 weeks post-implantation.
Answers are grounded in this paper and cite their sources.
Supports novel LV lead extraction feasibility in sheep; leaves open human safety data before clinical consideration.
Background A cardiac lead with a side helix for active fixation to the coronary vein wall (Attain Stability®, Model 20066, Medtronic, Minneapolis, MN, USA) recently received CE Mark. The lead is designed to improve left ventricular (LV) placement and reduce dislodgement rates. The extractability of this active fixation LV lead has not been studied extensively. Methods Seventeen sheep were implanted with either an LV lead with a side helix (Model 20066, Model 20096, Medtronic) or a unipolar LV lead (Model 4193, Medtronic) as a control. Leads were extracted at approximately 26, 52, or 118 weeks. Standard extraction methodology was employed with quantitative traction up to 907 g (2 lbs.) using a locking stylet. Gross pathology and histology of the heart with particular attention to the lead tracts were performed. Results All leads were successfully removed in their entirety and required significantly less than 1 kg of traction force. The side helix disengaged from the vein as designed and resulted in no complications. No cardiac tissue was observed on any extracted lead. Gross pathology and histology were devoid of any helix‐induced lesions in the vascular structures. The epicardium over the side helix was normal and the fibrotic reaction around the helix was not significantly different from that around the nonhelix portions of the study leads or the control lead. Conclusion Extraction of the side helix, active fixation LV lead from the coronary veins in the sheep model is safe, without procedural complexity, and free of complications after long‐term LV lead implant duration.
No takes yet. Share an insight, caveat, or question.
Adler et al. (2017) studied Left ventricular lead implantation and extraction (sheep model) (n=17). Left ventricular lead with a side helix for active fixation (Attain Stability Model 20066 and Model 20096) vs. Unipolar left ventricular lead (Model 4193) was evaluated on Complete lead extractability and extraction-related complications. Active fixation side-helix LV leads were extracted with 100% success and 0% complications at up to 118 weeks in 17 sheep, requiring <1 kg of traction force.
Synapse has enriched 10 related papers. Consider them for comparative context: